浊度仪说明书Read the specific instructions regarding calibration of your meter. Adjustments for calibrations will vary. Separate adjustments for low and high range should be used. 2Fill each of the test vials with distilled water and replace the caps. Place a piece of tape on ...
浊度仪说明书
Read the specific instructions regarding calibration of your meter. Adjustments for calibrations will vary. Separate adjustments for low and high range should be used. 2Fill each of the test vials with distilled water and replace the caps. Place a piece of tape on top of each cap. Clean the vials with tissue and glass cleaner. Turn on the turbidimeter and allow it to warm up for the required amount of time given by the instrument manufacturer. Place one of the vials in the meter and turn it to determine the orientation which gives the lowest NTU measure. Mark the tape on the cap with an arrow to indicate that orientation. 3Empty the vials and rinse with distilled water. In one of the vials, add the low turbidity standard to the 1/4-full mark. Gently swirl the vial and then dispose this solution. Add the standard solution to fill the vial and then insert it into the turbidimeter in the correct orientation. Read the value and adjust the meter to the correct reading if necessary, using the manufacturer instructions. Rinse the vial with distilled water afterwards. 4Add high turbidity standard to another vial to the 1/4-full mark. Gently swirl the vial and then dispose the solution. Fill the vial with the high turbidity standard and insert it into the meter. Read the turbidity and adjust the meter to obtain the correct reading in the high range. Dispose of the calibration solution and then rinse the vial with distilled water. Sponsored Links Gas Flow Calibrationwww.e.FlukeCal.com/CalibrationGuide Download a Free Best Practice Guide On Effective Gas Flow Calibration. Ekonogage by Hemcowww.hemcogages.com/ekonogage Pin gages, sets and libraries We've got the inventory so call us Prevent Pump Sand Damagewww.lakos.com LAKOS Sand Separators Submersible pumps and turbine pumps Confab Turbidimeterwww.confabinstrumentation.com Inline monitor measures turbidity 0-2 to 0-2000NTU in process line Tips & Warnings Choose a high calibration standard that is nearest to the maximum NTU value of the water being sampled. Check the calibration of the turbidimeter at least every other day. Calibration problems in the high range may be due to a worn light bulb. Calibration problems may occur if the battery charge is low. In-line process turbidimeter calibration can be adapted to this procedure if there are access ports for calibration and washing solution access. Rinsing between steps is a necessary procedure. Replace calibration standards before the expiration dates. Use the same cap on each vial to assure the proper orientation and that there is no cross-contamination of samples. Do not shake the vials when rinsing; this will introduce bubbles. Read more: How to Calibrate a Turbidimeter | eHow.com http://www.ehow.com/how_12043865_calibrate-turbidimeter.html#ixzz1j2hKfxTm
/
本文档为【浊度仪说明书】,请使用软件OFFICE或WPS软件打开。作品中的文字与图均可以修改和编辑, 图片更改请在作品中右键图片并更换,文字修改请直接点击文字进行修改,也可以新增和删除文档中的内容。
[版权声明] 本站所有资料为用户分享产生,若发现您的权利被侵害,请联系客服邮件isharekefu@iask.cn,我们尽快处理。 本作品所展示的图片、画像、字体、音乐的版权可能需版权方额外授权,请谨慎使用。 网站提供的党政主题相关内容(国旗、国徽、党徽..)目的在于配合国家政策宣传,仅限个人学习分享使用,禁止用于任何广告和商用目的。

历史搜索

    清空历史搜索